LifePak Defibrillator Specifications

The recorder can print the ECG trace, time and date, ECG lead and size, heart rate, and SYNC (if activated). The ac line power fuses and a ground connector are accessible at the rear panel. The LII~EPAK 9A defibrillator/monitor specifications are listed in Table 1—1.

Table 1-1 LIFEPAK 9A defibrillator/monitor Specifications*

ECG Monitor

 
ECG LEAD SELECTIONS

Std, Paddles, I, II, III

INPUT

Isolated ECG via QUIK-LOOK defibrillator paddles, FAST-PATCH disposable defibrillation/ECG electrodes, or 3-lead patient cable.

PATIENT CABLE LENGTH

Total length 4.0m (1311): cable 3.1m (101t), leads 0.9m (3ft).

COMMON MODE REJECTION

100d13 minimum with respect to chassis ground and 65dB minimum with respect to isolated ground when measured at 60Hz. Common mode range for patient cable input >—10V peak with respect to isolated ground.

MONITOR DISPLAY SIZE

102mm (4in) wide x 76mm (3in) tall, non-fade.

SWEEP SPEED

25mm/s

FREQUENCY RESPONSE

Non-diagnostic.

ECG Leads:

1.0 to 40Hz (—3dB)

Paddles:

2.2 to 20Hz (—3dB)

ECG SIZE

Adjusts amplitude of ECG trace on monitor and strip chart recorder.

HEART RATE METER

Three-digit readout displays rates from 20 to 300bpm. Heart rates outside this range do not yield valid systole tones or heart rate display.

1 mV Cal

Momentary pushbutton on rear panel simulates a l mV signal pulse to the ECG input.

Defibrillator

 
ENERGY SELECT

External Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50, 100, 200, 300, 360J
Internal Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50J

CHARGE CONTROLS

Independent momentary pushbutton controls on front panel and APEX paddle.

PADDLE CORD LENGTH

3m (10ft)

CHARGE INDICATORS

Flashing lamps on paddle and front panel pushbutton along with increasing stored energy display on monitor indicate charge in progress. Upon full charge, energy available is displayed and charge completed tone sounds.

*AII specifications at 20°C unless otherwise stated. Specifications subject to change without notice.