LifePak Defibrillator/Monitor Specifications

Table 1-1 LIFEPAK 9A defibrillator/monitor Specifications* (cont.)

CHARGE TIME

Charge to 360J in less than IOs with a fully-charged battery.

SYNC

Synchronizes defibrillator pulse to patient-generated QRS complex.

SYNC INDICATOR

Inverted triangle marker on displayed ECG waveform identifies synchronizer trigger point with respect to patient's QRS complex.

DEFIBRILLATOR ENERGY

+15% at 50Q

ACCURACY

± 30% over the range of 25 to 100Q

DEFIBRILLATOR WAVEFORM

5ms monophasic pulse (Edmark)

Thermal Array Recorder

PAPER
Size:

50mm wide x 30m (lOOft)

Speed:

25mm/s

DELAY

ECG prints 8 seconds after first appearing on the monitor.

FREQUENCY RESPONSE ECG

Non-diagnostic (diagnostic available via Setup Menu).

Leads (non-diagnostic): ECG

1.0 to 40Hz (-3dB). 0.05 to

Leads (diagnostic): Paddles:

100Hz (-3dB). 2.2 to 20Hz

ANNOTATION

(-3dB). Time, date, ECG lead, ECG gain, heart rate, defibrillation parameters, test load discharges.

 

*All specifications at 20°C unless otherwise stated. Specifications subject to change without notice.

Next Page

 