AC INPUT OPTIONS |
120 or 240Vac nominal line voltage without adjustment. 50 or 60Hz with adjustment in Setup Menu. |
BATTERY TYPE |
Sealed lead-acid, 16Vdc nominal, 3Ahr. |
BATTERY CAPACITY |
A new, fully-charged battery will provide one of the
following prior to shutdown:
|
Typical |
Minimum |
Number of 360J discharges |
75 |
40 |
Minutes of monitoring |
90 |
70 |
|
LOW BATTERY INDICATOR |
Advises operator to connect ac power. Battery capacities before and after warning
(LOW BATTERY–CONNECT AC POWER) for a new, fully charged battery are:
|
Typical |
Minimum |
Number of 360J discharges before warning |
50 |
20 |
Number of 360J discharges after warning |
25 |
5 |
|
BATTERY CHARGE INDICATOR
|
Indicator BATT CHRG illuminates when battery is charging |
BATTERY CHARGE TIME
|
24hrs to full capacity |
SERVICE INDICATOR
|
Message SERVICE indicates self-diagnostic routines have detected
improper operation requiring service attention |
POWER CONSUMPTION |
160W maximum while monitoring with recorder on and defibrillator charging |
SIZE |
|
Height: |
35.2 cm (13.9 in) |
Width: |
29.7 cm (11.7 in) |
Depth: |
31.0 cm (12.2 in) |
Weight: |
12.7 kg (28lbs) |
POWER CORD LENGTH |
3m (10ft.) |
STANDARD PADDLE ELECTRODE AREA |
82cm2 |
Environmental |
|
ATMOSPHERIC PRESSURE |
797 to 500mm Hg (-570 to 11,000ft) |
RELATIVE HUMIDITY |
0-95% (non condensing ) at 0- 34* C (32-94*F)
0-80% (non condensing ) at 35-45* C (95-113*F) |
TEMPERATURE RANGE |
0-45*C (32-113*F) operating
-30-65*C (-22-149*F) storage. |
*All specifications at 20°C unless otherwise stated.
Specifications subject to change without notice