LifePak Defibrillator/Monitor Specifications

Table 1—1 LIFEPAK 9A defibrillator/monitor Specifications*

AC INPUT OPTIONS

120 or 240Vac nominal line voltage without adjustment. 50 or 60Hz with adjustment in Setup Menu.

BATTERY TYPE

Sealed lead-acid, 16Vdc nominal, 3Ahr.

BATTERY CAPACITY

A new, fully-charged battery will provide one of the following prior to shutdown:

Typical
Minimum

Number of 360J discharges

75

40

Minutes of monitoring

90

70

LOW BATTERY INDICATOR

Advises operator to connect ac power. Battery capacities before and after warning (LOW BATTERY–CONNECT AC POWER) for a new, fully charged battery are:

Typical
Minimum

Number of 360J discharges before warning

50

20

Number of 360J discharges after warning

25

5

BATTERY CHARGE INDICATOR

Indicator BATT CHRG illuminates when battery is charging

BATTERY CHARGE TIME

24hrs to full capacity

SERVICE INDICATOR

Message SERVICE indicates self-diagnostic routines have detected improper operation requiring service attention

POWER CONSUMPTION

160W maximum while monitoring with recorder on and defibrillator charging

SIZE

 
Height:

35.2 cm (13.9 in)

Width:

29.7 cm (11.7 in)

Depth:

31.0 cm (12.2 in)

Weight:

12.7 kg (28lbs)

POWER CORD LENGTH

3m (10ft.)

STANDARD PADDLE ELECTRODE AREA

82cm2

Environmental
 
ATMOSPHERIC PRESSURE

797 to 500mm Hg (-570 to 11,000ft)

RELATIVE HUMIDITY

0-95% (non condensing ) at 0- 34* C (32-94*F)
0-80% (non condensing ) at 35-45* C (95-113*F)

TEMPERATURE RANGE

0-45*C (32-113*F) operating
-30-65*C (-22-149*F) storage.

*All specifications at 20°C unless otherwise stated. Specifications subject to change without notice