Item
|
Control or Indicator
|
Function
|
1
|
ON
|
Toggles power on and off. Power–on illuminates the LED. |
2
|
Lead Select
|
Selects PADDLES input or patient cable leads (STD, LEAD I, LEAD
II, or LEAD III). Switches selected lead display to reverse video and advances to the next
lead selection. Defaults to LEAD II at power–on unless PADDLES is specified in the Setup Menu. |
3
|
Energy Select
|
Displays the energy selection range incrementally. Select LOW on
the high range menu to display the low range selections from 0 to 9
joules. Select HIGH on the low range menu to display the
high range selection from 0 to 360 joules. Selecting OJ inhibits
energy charging. Pressing during charging or prior to energy
transfer causes stored energy to discharge internally. |
4
|
ECG Size
|
Changes the displayed ECG size to reverse video and increases or
decreases the gain by one setting. ECG size is adjustable from
0.2cm/mV to 4.0cm/mV. Defaults to gain of X1.0 at power–on. |
5
|
Charge
|
Initiates defibrillation charge cycle. Existing
defibrillator status message is replaced with XXX JOULES CHARGING message,
where XXX displays the stored energy as it ramps up to the selected value. LED
flashes during charge. At charge completion XXX JOULES AVAILABLE message
displays while LED remains on steadily. |
6
|
SYNC
|
Activates sync mode message display and LED light. LEI) flashes with each
QRS detection. Pressing SYNC again returns
to defibrillate (asynchronous) mode and turns off the LEI). |
7
|
Record
|
Toggles the recorder on and off. |
8
|
CRT Display
|
No-fade CRT display. ECG trace moves from right to left. |
9
|
QRS Indicator
|
Heart symbol flashes with each QRS detection. |
10
|
Heart Rate
|
Displays digital QRS rate (20 to 300bpm). |
11
|
Trace
|
No-fade display trace moves from right to left. |
12
|
Lead Select Indicator
|
Display indicates lead selected. |
13
|
Calibrated ECG Gain
|
Display indicates actual ECG gain selected. |