Item |
Control or Indicator |
Function |
14 |
Status Message Display |
Provides operating status information: |
SYNC MODE |
Indicates the synchronous mode is activated. |
FOR SYNC: USE LEADS |
Flashes if the SYNC pushbutton is pressed and the PADDLES
lead is selected with the standard paddles connected (the
synchronous mode may not be used with the PADDLES lead selected and standard paddles connected). |
NO PAPER |
Flashes if the recorder is activated but no paper is
installed. The recorder will not operate without paper. |
LOW BATTERY: CONNECT AC POWER |
Appears if the defibrillator/monitor is powered by battery and the battery voltage is low. As soon as this message appears, connect the defibrillator/monitor to ac power for continuous
operation and to recharge the battery. |
SERVICE |
Indicates the internal self-diagnostic software has detected a fault status condition. Refer
to page 3–39 for more information. |
15 |
Energy Selection Display |
Selection from high energy range shown. Press ENERGY
SELECT to display the high energy range menu. Select LOW
to display the low energy range menu. Energy selection range stays on screen for10 seconds following the last energy level selected or until pressing another pushbutton. |
16 |
JOULES selected Indicator |
On–screen indication of selected energy level in joules. Number increases up to selected
energy level as charging occurs. Defaults to 200J at
power–on. |
17 |
BATT CHRG |
Indicates battery is charging |
18 |
Recorder |
Records ECG and annotation in delay ECG mode only. Prints date, time, sync mode (if active),
ECG lead, ECG gain, and heart rate. Prints selected energy and defibrillation
mode during defibrillation. |
Paddles Area |
|
19 |
Patient Cable Connector |
Connection for 6-pin patient cable. |
20 |
Sternum Paddle |
Defibrillation electrode with one discharge pushbutton, usually placed to left of sternum
(patient's right). Also serves as negative ECG electrode during QUIK–LOOK
paddle monitoring. |