header2

Set-Up Menu-Operation

Set Up Menu

The Setup Menu allows you to examine and change various operating settings such as the language, notch filters, and the alarm volume. To access the Setup Menu, do the following:

1. First, access the Test Menu by turning power off. Then simultaneously press and hold LEAD SELECT and SYNC while pressing ON.

2. After the Test Menu appears as shown in Figure 2–3, press V ECG SIZE to scroll down to SETUP. Then press SYNC to display the Setup Menu as shown in Figure 2-4.

The different options are as follows:

Language

English, German, French, or Spanish

Recorder

Monitor or Diagnostic

Notch

50 Hz or 60 Hz

Leads Notch

Yes or No

Lead

Paddles or Lead II

Alarm Vol

1 - 8 (A short tone sounds with each level change to indicate the selected volume

3. Press SYNC to scroll through the selections. Press ¨ ECG SIZE A to scroll through the different option settings.

4. After you choose the desired option setting, press SYNC again to store the setting and move to the next selection. Whatever options are selected become the power-on default settings.

5. To exit the Setup Menu and return to the Test Menu, press LEAD SELECT. To exit the Test Menu and return to normal operation, press V ECG SIZE to scroll down to EXIT, then press SYNC.