Information Menu-OperationInfo MenuThe Info Menu allows you to examine the manufacturing date and EPROM revision numbers, and to examine and clear the number of defibrillator discharges recorded and fault status codes recorded. To access the Info Menu, do the following: 1. First, access the Test Menu. Turn power off, then simultaneously press and hold LEAD SELECT and SYNC while pressing ON. 2. After the Test Menu appears as shown in Figure 2—3, press V ECG SIZE to scroll down to INFO. Then press SYNC to display the Info Menu as shown in Figure 2—5.
1. Press SYNC to scroll through the selections. For DEFIB USES: or FAULT STATUS:, press V ECG SIZE A to toggle between CLEAR and RETAIN. 2. To exit the Info Menu and return to the Test Menu, press LEAD SELECT. To exit the Test Menu and return to normal operation, press V ECG SIZE to scroll down to EXIT, then press SYNC. |