LifePak PIP-Trouble Shooting

PIP-Discharge Control

1. Remove the paddles from the paddle storage area. Extend the coil cords away from the paddle wells at least three feet and away from your body or any other object.

WARNING:

SHOCK HAZARD. Help avoid injury by holding the paddles and coil cords away from your body and from any person or object as described in the following steps.

2. Press CHARGE.

3. After the defibrillator reaches full charge (indicated by the tone and the displayed message 360 JOULES AVAILABLE) keep the coil cords extended and away from your body or any other object while shaking the paddles and coil cords with a moderate motion. Confirm that no discharge occurs.

4. Return the paddles to the paddle storage area.

5. Simultaneously press both discharge pushbuttons and confirm that the energy is discharged into the test load.

PIP Energy Dump

1. Place the paddles on the energy meter electrodes.

2. Press A ENERGY SELECT V and set the energy level to 360J.

3. Press CHARGE.

4. After the defibrillator/monitor charges to 360J, change the energy level to 300J.

5. Confirm CHARGE REMOVED is displayed and tones sound. Return the paddles to the paddle storage area.

PIP Charge Reduction

1. Leave the paddles stored in the paddle storage area. Press A ENERGY SELECT V and set the energy level to 360J.

2. Press CHARGE.

3.  When full charge is reached, start the timer and immediately change the energy level to lJ and press CHARGE (to perform this test properly, you should scroll to 1J and press CHARGE within a few seconds).

4. Confirm that full charge of 1J is reached within 20 seconds or less after you start the timer. Simultaneously press both discharge pushbuttons and confirm that the energy is discharged into the test load.