LifePak Maximum Leakage-Trouble Shooting

WARNING
ELECTRICAL HAZARD. Make sure you are familiar with your
test equipment and how to execute these tests. Failure to properly
perform these tests could result in not detecting excessive leakage
current.
TABLE 3-3 Maximum Leakage Current for 120Vac and 60Hz

NOTE:
1. All tests are conducted in both hot/neutral normal
and reverse polarities.
2. Test procedure limits based on AAMI/ANSI Standards
with meter inaccuracies factored in.
3. N/A = Not Applicable.
4. Ground resistance: <0.050Q as measured between the
rear panel ground stud and the ground pin on the ac power connector
(refer to Figure 3-6).
Next Page
|