LifePak PIP-Trouble Shooting

PIP-Scope and Applicability

This PIP applies to all configurations of the LIFEPAK 9A defibrillator/ monitor. This PIP does not apply to the LIFEPAK 9, LIFEPAK 9B, or LIFE;PAK 9P defibrillator/monitors; refer to their respective service manuals for testing information

PIP Definitions
Acronyms used in this procedure are:
BPM

Beats Per Minute

ECG

Electrocardiogram

PIP

Performance Inspection Procedure

QRS or QRS Complex

Refers to portions of the ECG waveform

NSR

Normal Sinus Rhythm

p-p

peak-to-peak

NOTE:

Pushbuttons are indicated in CAPITAL LETTERS and displayed messages are indicated in BOLD CAPITAL LETTERS.

PIP Requirements

The requirements for test equipment, workstation, and personnel are listed below along with important safety information.

PIP Required Equipment

Perlbrming this PIP requires the equipment listed in Table 3-1. Although specific test instruments are recommended, other test equipment with equivalent specifications may he used.

Table 3-1 PIP Test Equipment

Equipment
Specifications
Manufacturer
Defibrillator

Power Range: 0-400J

Bio-Tek QED-6 or equivalent*

Energy Meter

Load Resistance: 5052 ±1%
Accuracy: ±3%
Defib Waveform Amplitude
Attenuation: 1000:1 (minimum)
with oscilloscope output

 
ECG simulator

Accuracy: 1% of selected rate

Bio-Tek QED-6 or equivalent*

Leakage Tester/
Safety Analyzer

11OVac Line Voltage:
Current Range: 0-2000µA
Current Accuracy: ±1%
Resistance Accuracy: ±1%

Dynatech Nevada 232D, or equivalent*

230Vac Line Voltage:
Current Range/Accuracy:
200µA/5% of reading
2000µA/l% of reading
Resistance Accuracy: ±1%

Dynatech Nevada 432 HD, or equivalent*

* Equivalent equipment is required to meet the specifications listed in the specifications column.