LifePak Calibration-Trouble Shooting

Table 3-5 Test and Calibration Test Equipment

Equipment
Specifications
Manufacturer
Storage Oscilloscope

Bandwidth: dc to 2kHz
Vertical Accuracy: -15% (5mV-5V/division)
Horizontal Time Base Accuracy: ±1%

Tektronix Model 2232 or
equivalent*

Digital Multimeter (DMM)

4-1/2 digits, 2A Range
DC Voltage Accuracy: 0.1% of reading
Low Resistance: 0.01Q resolution on 20052 scale

Fluke Model 8012A or equivalent*

Function Generator

Output Waveform: Sinusoidal
Amplitude/Freq. Accuracy: ±0.5dB/±5%
Amplitude/Freq. Range: 10V/11MHz

Krohn-Hite Model 5400A or equivalent*

ECG Simulator

Sinewave Output: 1.0 ±0.5mV p-p
Output Frequency: 10.0 ±0.5Hz

Bio-Tek QED-6 or equivalent*
Defibrillator Energy Meter

Power Range: 0-40W
Load Resistance: 5052±1%
Accuracy: ±3%
Defib Waveform Amplitude Attenuation:
1000:1 (min.) with oscilloscope output

Bio-Tek QED-6 or equivalent*
DC Power Supply

Output Voltage: 0 to 20Vdc
Output Current: 0 to 3A

Topward Model 'I'PS 2000 or equivalent*

Timer
Resolution: 1 second Aristo or equivalent*
Components for Recorder
Frequency Response test
(construction defined in
Figure 3—13, page 3—35)
R1, R2 Resistor, 100k 1/8W, 1%
R3, R4 Resistor, 100 1/8W, 1%

Physio-Control PN 200054-384
Physio-Control PN 200054-097

Phone plug
0.25  inch plug tip  

* Equivalent equipment is required to meet the specifications listed in the specifications column.

TCP-Test Equipment Verification

All test equipment used in performing the TCP must have a current calibration label affixed to its exterior. The calibration label must be issued by a certified calibration facility.

TCP-Workstation Requirements

Workstations where the TCP is performed must provide an Electrical Static Discharge (ESD) protected environment. The ac line power for this workstation must be connected to a grounded power source.

TCP-Personnel

Personnel performing this TCP must be thoroughly familiar with the operation of the LIFEPAK 9A defibrillator/monitor. In addition, personnel performing this TCP must meet at least one of the following levels of education or experience:

1. Associate of applied science degree with a major emphasis in biomedical electronic

2. Certificate of technical training in electronics with a major emphasis in biomedical electronics

3. Equivalent biomedical electronics experience.