LifePak Calibration-Trouble Shooting
Table 3-5 Test and Calibration Test Equipment
Equipment
|
Specifications
|
Manufacturer
|
Storage Oscilloscope
|
Bandwidth: dc to 2kHz
Vertical Accuracy: -15% (5mV-5V/division)
Horizontal Time Base Accuracy: ±1% |
Tektronix Model 2232 or
equivalent* |
Digital Multimeter (DMM)
|
4-1/2 digits, 2A Range
DC Voltage Accuracy: 0.1% of reading
Low Resistance: 0.01Q resolution on 20052 scale |
Fluke Model 8012A or
equivalent* |
Function Generator
|
Output Waveform: Sinusoidal
Amplitude/Freq. Accuracy: ±0.5dB/±5%
Amplitude/Freq. Range: 10V/11MHz |
Krohn-Hite Model 5400A or equivalent* |
ECG Simulator
|
Sinewave Output: 1.0 ±0.5mV p-p
Output Frequency: 10.0 ±0.5Hz |
Bio-Tek QED-6 or equivalent* |
Defibrillator Energy Meter
|
Power Range: 0-40W
Load Resistance: 5052±1%
Accuracy: ±3%
Defib Waveform Amplitude Attenuation:
1000:1 (min.) with oscilloscope output |
Bio-Tek QED-6 or equivalent* |
DC Power Supply
|
Output Voltage: 0 to 20Vdc
Output Current: 0 to 3A |
Topward Model 'I'PS 2000 or equivalent* |
Timer
|
Resolution: 1 second |
Aristo or equivalent* |
Components for Recorder
Frequency Response test
(construction defined in
Figure 3—13, page 3—35)
|
R1, R2 Resistor, 100k 1/8W, 1%
R3, R4 Resistor, 100 1/8W, 1% |
Physio-Control PN 200054-384
Physio-Control PN 200054-097 |
Phone plug
|
0.25 inch plug tip |
|
* Equivalent equipment is required to meet the specifications listed
in the specifications column.
TCP-Test Equipment Verification
All test equipment used in performing the TCP must have a current
calibration label affixed to its exterior. The calibration label must be
issued by a certified calibration facility.
TCP-Workstation Requirements
Workstations where the TCP is performed must provide an Electrical
Static Discharge (ESD) protected environment. The ac line power for this
workstation must be connected to a grounded power source.
TCP-Personnel
Personnel performing this TCP must be thoroughly familiar with the
operation of the LIFEPAK 9A defibrillator/monitor. In addition,
personnel performing this TCP must meet at least one of the following
levels of education or experience:
1. Associate of applied science degree with a major emphasis in
biomedical electronic
2. Certificate of technical training in electronics with a major
emphasis in biomedical electronics
3. Equivalent biomedical electronics experience.