TCP Test Menu-Trouble Shooting

TCP-Test Menu Access

Three of the four tests on the Test Menu (DEFIB, DISPLAY, and Access RECORDER) are used in Test and Calibration Procedures. To access the Test Menu, do the following:

1. Turn the defibrillator/monitor power off.

2. Press ON while pressing and holding down LEAD SELECT and SYNC until the Test Menu is displayed as shown in Figure 3—8.

Press ECG SIZE to scroll through the selections.

Press SYNC to start a test.

Press LEAD SELECT to interrupt a test in progress and return to the Test Menu.

To exit the test menu, press ECG SIZE to scroll to EXIT, then press SYNC.

TCP-Defibrillator Calibration

1. With the defibrillator/monitor power off, place the paddles on the energy meter electrodes.

CAUTION

Possible Energy Meter Damage. Start each test at the lowest energy level to avoid overheating the energy meter. Test each energy level in successive order.

2. Access the Test Menu (turn power off, then press ON while pressing and holding down LEAD SELECT and SYNC until the Test Menu is displayed).