LifePak TCP-Trouble Shooting

TCP - Test Load Calibration

NOTE:

Complete TCP-Defibrillator Calibration before performing this calibration of the test load.

1. Make sure the paddles are securely stored in the paddle storage area of the defibrillator.

2. Access the Test Menu (turn power off, then press ON while pressing and holding down LEAD SELECT and SYNC until the Test Menu is displayed).

3. Press ECG SIZE to scroll to the DEFIB test.

4. Press SYNC to start the DEFIB test.

5. Press V ENERGY SELECT A to set the selected energy to 200J.

6. Press CHARGE.

7. After the defibrillator reaches full charge, simultaneously press the paddle discharge pushbuttons to transfer the energy into the test load.

8. When the CRT displays JOULES DELIVERED: 200, press SYNC.

9. Press LEAD SELECT to return to the Test Menu.

10. Scroll to EXIT and press SYNC to resume normal operation. Select STD lead.

11. Press CHARGE. After the defibrillator reaches full charge, simultaneously press the paddle discharge pushbuttons to transfer the energy into the test load.

12. Confirm that the CRT displays TEST 200 JOULES DELIVERED.

13. Confirm that the recorder prints the time and date followed by TEST 200 JOULES DELIVERED.