LifePak TCP-Trouble Shooting

TCP - Display Calibration
The following paragraphs describe the calibration
procedure for the Display Assembly PCB. Figure 3—10 shows the location
of all Display Assembly PCB potentiometers.
1. Disconnect the ac power cord. Remove front and rear
cases to allow access to the potentiometers. For disassembly
instructions, refer to page 4—8, Case Separation.
WARNING
SHOCK HAZARD. The Display Assembly produces high
voltages up to +11.6kV. Do Not contact the CRT anode connector or the
CRT bias generator circuitry when the defibrillator/monitor power is on.
WARNING
FLYING GLASS HAZARD. The CRT is made of glass and contains a
vacuum which may implode if broken. Handle with care and wear safety
glasses when removing the CRT from the Display Assembly PCB.
2. Reconnect the ac input cord. Reconnect J1 on the front panel.
NOTE:
Reroute Jl ribbon cable underneath C32 on the Power Conversion PCB to
prevent strain on the capacitor during the following tests.