LifePak PIP-Trouble Shooting

Table 3-1 PIP Test Equipment (cont)

Equipment
Specifications
Manufacturer
Patient Cable
3-lead, AHA
or 3-lead IEC
Physio-Control PN 9-10418-02
Physio-Control PN 800947-01
Recorder Paper
50mm (use only Physio-Control paper) Physio-Control PN 804700M
Metric Ruler or
recorder paper
Resolution: 1mm divisions  
Timer
1 second resolution Aristo or equivalent*

* Equivalent equipment is required to meet the specifications listed in the specifications column.

CAUTION

Possible Equipment Damage. Be sure to use only recorder paper from Physio-Control (PN 804700). Other paper may damage the printhead and may not print legibly.

PIP - Test Equipment Verification

All test equipment used in performing the PIP must have a current calibration label affixed to its exterior. The calibration label must be issued by a certified calibration facility.

PIP Workstation

The ac line power for this workstation must be connected to a grounded power source.

PIP Personnel

Personnel performing this PIP must be thoroughly familiar with the operation of the LIFEPAK 9 defibrillator/monitor. In addition, personnel performing this PIP must meet at least one of the following levels of education or experience:

1. Associate of applied science degree with a major emphasis in biomedical electronics

2. Certificate of technical training in electronics with a major emphasis in biomedical electronics

3. Equivalent biomedical electronics experience

WARNING

SHOCK HAZARD. When charged and discharged during this PIP, the LIFEPAK 9 defibrillator/monitor discharges up to 36W of electrical energy through the defibrillator paddles. Unless discharged properly as described in this PIP, this electrical energy may cause injury or death. Do not attempt to perform this procedure unless you are thoroughly familiar with the operation of the LIFEPAK 9 defibrillator/monitor.

Next Page