LifePak PIP-Trouble ShootingTable 3-1 PIP Test Equipment (cont)
* Equivalent equipment is required to meet the specifications listed in the specifications column. CAUTIONPossible Equipment Damage. Be sure to use only recorder paper from Physio-Control (PN 804700). Other paper may damage the printhead and may not print legibly. PIP - Test Equipment VerificationAll test equipment used in performing the PIP must have a current calibration label affixed to its exterior. The calibration label must be issued by a certified calibration facility. PIP WorkstationThe ac line power for this workstation must be connected to a grounded power source. PIP PersonnelPersonnel performing this PIP must be thoroughly familiar with the operation of the LIFEPAK 9 defibrillator/monitor. In addition, personnel performing this PIP must meet at least one of the following levels of education or experience: 1. Associate of applied science degree with a major emphasis in biomedical electronics 2. Certificate of technical training in electronics with a major emphasis in biomedical electronics 3. Equivalent biomedical electronics experience WARNINGSHOCK HAZARD. When charged and discharged during this PIP, the LIFEPAK 9 defibrillator/monitor discharges up to 36W of electrical energy through the defibrillator paddles. Unless discharged properly as described in this PIP, this electrical energy may cause injury or death. Do not attempt to perform this procedure unless you are thoroughly familiar with the operation of the LIFEPAK 9 defibrillator/monitor. |