TCP-Trouble Shooting

2. Reconnect the battery to the defibrillator/monitor.

NOTE:

The Power Supply compensates for temperature fluctuations. The defibrillator/monitor must be powered—on for at least 30 minutes prior to calibration so the temperature around thermistor RT2 stabilizes.

3. Connect the DMM between the positive and negative battery terminals.

4. Connect the ac power cord and press ON.

5. Locate R23. Remove the potting compound from R23 as required to allow adjustment. Adjust R23 for +18.40 ±0.18V.

TCP - Recorder Calibration

1. Remove the front panel. The two potentiometers (R176 and R8 on the Main PCB) Required for recorder Calibration are accessible with the front panel removed as shown in Figure 3-12.