TCP-CRT-Trouble Shooting

TCP - CRT Frequency Response

1. Press ECG SIZE to increase the ECG gain to X1.4.

2. Connect the function generator to the patient cable through the 1000:1 divider (as illustrated in Figure 3–13
on page 3–35).

3. Set the function generator for a lVp-p, 10Hz sinewave.

4. Decrease the function generator signal frequency to 1 Hz. Confirm that the trace displayed on the CRT is
19 ±2mm p-p.

5. Increase the function generator signal frequency to 40Hz. Confirm that the trace displayed on the CRT is
19 ±2mm p-p.

6. Disconnect the function generator.