LifePak PIP-Trouble Shooting

PIP Battery Charge
With defibrillator/monitor power on and the ac power source connected
confirm that BATT CHRG is displayed in the lower left corner of
the display.
PIP-Battery Power
1. With defibrillator/monitor power on, disconnect the power cord
from the ac power source. Confirm that the display remains on but
BATT CHRG is not displayed.
2. Reconnect the ac power cord and press ON to turn off
defibrillator/monitor power.
PIP-Setup Menu Settings
This procedure describes how to access the Setup Menu and select the
proper settings required for the rest of the PIP. Before changing any
settings, write down the current settings so you can restore these
settings at the completion of the PIP.
1. To access the Setup Menu, make sure defibrillator/monitor power is
off. Simultaneously press and hold LEAD SELECT and SYNC, then press ON.
2. Continue holding LEAD SELECT and SYNC until the message
SELECTED TEST: and a list of tests are displayed as shown in Figure
3-2. This information is the Test Menu.
Next Page
|