LifePak PIP-Trouble Shooting
PIP-Keypad Test
1. The Keypad Test is initiated from the Test Menu (as shown in
Figure 3-2) which should be displayed from the end of the previous step
Press ECG Size to scroll to KEYPAD, then press SYNC to start the Keypad
Test.
2. Press teh front panel pushbuttons and then the rear panel
pushbuttons in the following order:
FRONT PANEL:
1. LEAD SELECT
2. ECG SIZE (down arrow)
3. ECG SIZE (up arrow)
4. ENERGY SELECT (down arrow)
5. ENERGY SELECT (up arrow)
6. CHARGE
7. RECORD
8. SYNC
REAR PANEL:
1. QRS VOL (down arrow)
2. QRS VOL (up arrow)
3. MODE
4. 1mV CAL
5. SET
3. Press the APEX paddle CHARGE pushbutton
4. Press the APEX paddle discharge pushbutton; confirm no beep
sounds.
5. Press the STERNUM paddle
discharge pushbutton; confirm no beep sounds.
6. Simultaneously press both paddle discharge pushbuttons; confirm
that a beep sounds.
7. Confirm that the monitor displays KEYPAD: OK, then press
LEAD SELECT to return to the Test Menu. If the pushbuttons are not
pressed in the order listed above, or if there is a pushbutton failure,
the monitor displays KEYPAD: FAULT. In this case, press LEAD
SELECT to return to the Test Menu; press V ECG SIZE A to
scroll to KEYPAD, then press SYNC to restart the Keypad Test and
repeat steps 2 through 7.
8. To exit the Keypad Test after completion, press ECG SIZE to scroll
to EXIT and press SYNC to resume normal operation.