LifePak PIP-Trouble ShootingPIP - Lead Select1. Use the patient cable to connect the ECG simulator to the
defibrillator/monitor. Set the ECG simulator for a 2. Press LEAD SELECT repeatedly and confirm that the signal displays in all lead (I, II, III) configurations. PIP - Recorder1. Confirm there is paper in the recorder. 2. Set the ECG simulator for a 60bpm-NSR signal in LEAD II. 3. Press RECORD, let the recorder run for approximately 10 seconds, then select STD. Let the recorder run for an additional 10 seconds, then press RECORD to stop the recorder. 4. Measure the distance between the maximum peaks of the recorded ECG trace. Confirm that the distance between consecutive peaks is 25 +/- 1mm. 5. Measure the baseline centering of the recorded STD trace. Confirm that the baseline is centered within +/- 1mm. PIP - ECG DisplayConfirm that the flatline display on the CRT appears level and centered. Confirm that all messages are clearly visible as shown in Figure 3-4
|