LifePak PIP-Trouble ShootingPIP - 1mV CALIn order to perform this test the LEADS NOTCH must be set to NO and the recorder must be set to DIAG as described in the PIP-Setup Menu Settings. 1. Confirm that the STD lead remains selected from the previous PIP test (press LEAD SELECT if necessary to select STD). 2. Press RECORD, then press the 1 mV CAL pushbutton on the rear panel once. Confirm that a 10.0 ±0.5mm p-p calibration pulse is printed out as illustrated in Figure 3–5.
PIP - CRT Gain/Aspect Ratio1. Select LEAD II. Ratio 2. Set the ECG simulator for a 60bpm-NSR signal. 3. Adjust ECG SIZE for a 34 ±2mm p-p ECG display signal. 4. Press RECORD. 5. Confirm that the signal amplitude printed on the ECG paper is 20±1mm p-p. PIP - Heart Rate1. With LEAD II selected, set ECG Size to X1.0. 2. Set the ECG simulator for the input rates listed below. Confirm that the systole indicator (the heart-shaped symbol) flashes on the display with each QRS complex and that the displayed HR corresponds:
3. Return the ECG Simulator selection to 60 bpm-NSR. |